These converters feature low leakage current and a reinforced insulation system – critical for medical equipment applications – rated for 300 VAC working voltage and 4,200 VACrms input/output isolation. They comply with the UL/CSA/EN/IEC 60601-1 medical safety standard and are rated for 2X means of operator protection (MOOP).
AEE10W-M series single and dual output converters cover input voltages from 9 to 75 Vdc, with three 2:1 input voltage ranges of 9–18 V, 18–36 V and 36–75 V. Each of the three input voltage categories offers a choice of two models, which produce a fully regulated 5 V or 12 V output. Rated at 10 watts, the converters have an operating temperature range of -40 to 85˚C, with full power available up to 60˚C. Typical efficiency is 82%.
The converters are fully encapsulated in an industry-standard 2 x1 inch (50.8 x 25.4 mm) low profile aluminum case. Standard features include latching overvoltage and overtemperature protection, with hiccup mode overcurrent and short-circuit protection. All AEE10W-M series converters use 150 kHz fixed frequency switching and comply with EN 55022 class A and FCC level A standards for conducted and radiated EMI.
-
Total Max Power:
10 W
-
Total Max Current:
1.6 A
-
Input Voltage:
9 V to 75 V
-
Output Voltage:
-15 V to 15 V
-
Outputs:
Single, Dual
-
Size (L x W x H):
2.000" x 1.000" x 0.400"
-
Warranty:
3 years
-
Operating Temperature:
-40 to +75 degC
-
- 4200VAC reinforced Insulation
-
- Insulation rated for 300VAC Working Voltage
-
- Medical Safety to UL/CSA/EN/IEC
-
- 60601-1 3rd Edition
-
- 2 MOOP rated
-
- Wide 2:1 Input Voltage Range
-
- Fully regulated Output Voltage
-
- Low Leakage Current
-
- Operating Temp. Range –40°C to 75 °C
-
- Input Filter meets EN 55022, class A and FCC, level A
-
- Overload Protection
-
- 2”x 1” Plastic Package
-
- 3 Years Product Warranty