These converters feature low leakage current and a reinforced insulation system – critical for medical equipment applications – rated for 300 Vac working voltage and 4,200 Vacrms input/output isolation. They comply with the UL/CSA/EN/IEC 60601-1 medical safety standard and are rated for 2X means of patient protection (MOPP).
AEE20W-M series single and dual output converters cover input voltages from 9 to 75 Vdc, with three 2:1 input voltage ranges of 9–18 V, 18–36 V and 36–75 V. Each of the three input voltage categories offers a choice of six models, which produce either single outputs of 5 V, 12 V, 15 V and 24V, or dual outputs of +/-12 V and +/-15V. Rated at 20 watts, the converters have an operating temperature range of -40 to 85˚C, with full power available up to 60˚C. Typical efficiency is 89% maximum.
The converters are fully encapsulated in an industry-standard 2 x1 inch (50.8 x 25.4 mm) low profile plastic case. Standard features include latching overvoltage and overtemperature protection, with hiccup mode overcurrent and short-circuit protection. All AEE20W-M series converters use 285 kHz fixed frequency switching and comply with EN 55022 class A and FCC level A standards for conducted and radiated EMI.
-
Total Max Power:
20 W
-
Total Max Current:
4 A
-
Input Voltage:
9 V to 75 V
-
Output Voltage:
-15 V to 24 V
-
Outputs:
Single, Dual
-
Size (L x W x H):
2.000" x 1.000" x 0.470"
-
Warranty:
2 years
-
Operating Temperature:
-40°C to 85°C
-
4200 Vac reinforced insulation
-
Insulation rated for 300 Vac working voltage
-
Medical Safety to UL/CSA/EN/IEC
-
60601-1 3rd Edition
-
2 MOPP rated
-
Wide 2:1 input voltage range
-
Fully regulated output voltage
|
-
Low leakage current
-
Operating temp. range –40°C to 75 °C
-
Input filter meets EN 55022, class A and FCC, level A
-
Medical EMC Standard meets 4th Edition of EMI EN55011 and EMS EN60601-1-2
-
Overload protection
-
2”x 1” plastic package
-
2 years product warranty
|