Normal operation conditions:Battery
voltage:in range of “BATT.GOOD”
-
Response time:within
3 sec. (It may time in measuring a capacitance loading)
-
Insulation resistance:≧100MΩ / DC1000V between test leads and enclosure.
-
Withstand voltage:AC5500V
(50 / 60Hz) for 1 min. between test leads and enclosure.
-
Overload Protection:between
test leads (for 10 sec.)
-
Battery:AA 1.5V
x 4
-
Safety standard:IEC
61010-1 CAT.III 600V IEC 61557
Measurement Ranges:
AC Voltage Range:
Voltage at 50 / 60 Hz
|
ACV 0~600V
|
Accuracy
|
Within ±3% of full scale
|
Insulation Resistance Range:
MODEL DE-5011 23°C ± 5°C max. 80%R.H
Nominal Voltage
|
125V
|
250V
|
500V
|
Max. Effective Scale Value
|
20 MΩ
|
50 MΩ
|
100 MΩ
|
Mid-Scale Value
|
0.5MΩ
|
1MΩ
|
2MΩ
|
Primary Effective Measuring Ranges
& Accuracy
|
0.02~10 MΩ
|
0.05~20 MΩ
|
0.1~50 MΩ
|
Within ±5% of indicates value
|
Secondary Effective Measuring
Ranges & Accuracy
|
Scale out of primary effective
measuring range (except 0, ∞)
|
Within ±10% of indicates value
|
Accuracy at 0 & ∞
|
Within ±0.7% scale length from 0 to ∞
|
Accuracy at no-load voltage
|
0%~+20% of nominal voltage
|
Nominal Current
|
1mA 0%~+20%
|
Short-circuit Current
|
Within 1.5mA
|
*Coarse-line arc on scales indicate
the primary effective measuring ranges. (guaranteed accuracy range)
TEMP./Humidity:
-
Operation:0°C~ 40°C(32℉~104℉)/<85%
R.H.(no condensation)
-
Storage:-20°C~ 60°C(-4℉~140℉)/<75%R.H.(no
condensation)
Dimension & Weight:
-
162(L)x112(W)x50(H)mm
-
Approx.400g (Excluding battery)
Accessories:
-
LINE test lead x 1
-
EARTH test lead plug x 1
-
Convertible test lead x 1
-
Alligator clip x 1
-
Carrying case x 1
-
Battery AA 1.5V x 4
-
Instruction Manual x 1